Introduction
Principle
Reagent Storage, Stability & Handling
AutoPure Bilirubin is a ready to use two liquid reagent system.
Shelf life
Stable till the expiry date indicated on the label, when stored at 2- 8 C.
Preparation of Working Reagent
On Board Reagent Stability
Note: Over a period of time, the total and Direct bilirubun working reagents may develop a yellow tinge. This is expected and does not impair performance of the assay.
Specimen collection & Preservation
Collect sample using standard sampling tube. Although serum is preferred, heparinized plasma or EDTA plasma can be used . Haemolysis interferes. Store samples protected from light as bilirubin is light sensitiv. Bilirubin in serum/plasma is stable for 1 day at 2-8 C and for 30 days at -20 C. Centrifuge samples containing precipitate before performing assay .
Calculations
Application Sheet
Refer to the application sheet for details. For additional system applications, contact our local accurex representative.
Calibration
For calibration , it is recommended to use C.F a.s or any other suitable calibrator material.
Calibration frequency
Re- calibration is recommended
Quality Control
Each batch of autopure bilirubin is assayed with atleast six quality control sera- Precinorm, precipath, bioradl, bioradll, accutrol normal and Accutrol abnormal prior to release.
To ensure adequate quality control, it is recommended that the laboratory should use a normal and abnormal commerical reference control serum. It should be realized that the use of quality control material checks both the reagent and instrument functions together.
If the control values fall outside the specified limits, each of the below criteria should be cross checked and corrected:
Expected Values
Serum/Plasma
Total bilirubin- Adults : 0.1-1.1 mg/dl( 1.71-18.8 mmol/l)
Infants 1.2 - 12 mg/dl( 20.52 -205.2 mmol/l)
Note:
Expected range varies from population to population. it is therefore recommended that each laboratory should establish its own normal range. For diagnostic purposes, the bilirubin results should always be assessed in diagnostic purposes, the bilirubin results should always be assessed in conjuction with the patients medical history clinical examinations and other findings.
Performance Characteristics
Linearity
With Autopure Bilirubin , the assay is linear up to 20 mg/dl (342 mmol/l).Determine samples with higher concentration via the rerun function . On instrument without rerun function, manually dilute samples with higher concentration using 0.9% NaCl or distilled/deionized water (e.g. 1+3 ). Multiply the result by the appropriate dilution factor (e.g. 4).
Interference
Haemolysis interferes in samples from adults. Occasionally , slighty higher values for direct bilirubin than for total bilirubin may be yielded by samples in which the totl and direct bilirubin concentrations are similar. In such samples, the bilirubin concentration determined by both methods should be stated.
Co- Relation Studies
A comparison of the bilirubin determination using autopure bilirubin and reference competitors bilirubin reagent gave the following co- realtion (mg/dl):
Linear Regression
Total
y= 0.974x +0.101
r= 0.9995
Sy.x= 0.22
No of samples measured :15
The samples concentrations measured were between:
Total :0.70 mg/dl and 20.0 mg/dl
Direct : 0.40 mg/dl and 8.70 mg/dl
References
![]() |
ACCUREX BIOMEDICAL PVT. LTD.
All Rights Reserved.(Terms of Use) Developed and Managed by Infocom Network Private Limited. |